UT-ONE Image Loading...
UT-ONE L01A
UT-ONE Image Loading...
UT-ONE S01A
UT-ONE Image Loading...
UT-ONE B03B
UT-ONE Image Loading...
UT-ONE S04A
UT-ONE Image Loading...
UT-ONE S04B
UT-ONE Image Loading...
UT-ONE S12A
UT-ONE Image Loading...
UT-ONE S12B
UT-ONE Image Loading...
UT-ONE S16B
UT-ONE Image Loading...
Comparison Table
Batemika UT-ONE S16B 16-channel Thermometer Readout
UT-ONE S16B
UT-ONE S16B
UT-ONE S16B
UT-ONE S16B
UT-ONE S16B

UT-ONE S16B is an accurate and versatile 16-channel benchtop thermometer readout. UT-ONE S16B is capable of measuring platinum resistance thermometers, thermistors and thermocouples. It has a large internal memory for storing measured data (at least 30 days of continuous acquisition). The user interface is realized using the acquisition software on a personal computer.

Resistance probes (PRTs) can be measured in 4-wire, 3-wire or 2-wire connection configuration. Banana connectors are user friendly, easily used with banana plugs, bare wires or spade lugs.

UT-ONE S16B can be connected to the computer via the RS232 or USB interface using the UT-ONE Applications software. Power to thermometer readout is supplied via USB cable.

Technical data
Main channels:
16 (consecutive)
Acquisition time per channel:
1 to 240 s
Keep-warm currents:
Automatically set on inactive channels
Probe Types:
Platinum resistance thermometers, Thermistors, Thermocouples
Temperature range:
-273 °C to 2100 °C max, limited by the thermometer probe
Cold junction compensation:
Internal, External
Communication interface:
USB (standard), RS232 (optional), GPIB (optional)
USB port power supply:
5 V DC, 0.3 A max
Power consumption:
0.75 W typical, 1.5 W max
Weight:
3.1 kg
Ext. dimensions (WxHxD):
500x90x155 mm
Ambient conditions monitoring:
Optional external probe for air temperature & relative humidity

More information is available in UT-ONE S16B User ManualPdf_icon Loading....

Specification summary
Probe: Platinum resistance th. Thermistors Thermocouple
Probe type: Pt100, Pt25, Pt200 NTC 2k, 3k, 10k K, J, T, E, N, S, R, B
Characterization: CVD, Polynom, ITS-90 SteinHart-Hart, Polynom Deviation function
Accuracy (Typical): ±0.004 °C ±0.001°C ±0.1 °C (Ext. CJC)
Effective resolution: 0.0005 °C 0.0001 °C 0.01 °C
Measurement current: 1 mA, 0.707 mA ±1% 10 μA, 7.07 μA ±1% n/a

If you would like to compare between thermometer specifications, please visit: Comparison Table

More information is available in UT-ONE S16B User ManualPdf_icon Loading....

Order information
Description
Part no.
Prices*
UT-ONE S16B 16-channel Thermometer Readout
6390.00 €
Ambient conditions probe
250.00 €
Plastic carrying case
250.00 €
R210A Reference resistor
590.00 €
LabVIEW driver and application
**

* Stated price is recommended price for customers in EU. Send and inquiry to info@batemika.com to get a formal quote. Price is given excluding VAT and shipping/handling costs. Batemika reserves the right to change stated prices without prior notice.

** Available for free-of-charge download at www.batemika.com

UT-ONE S16B Thermometer Readout Photo Gallery
UT-ONE S16B Gold-plated banana conectors
Gold-plated banana plugs - Pt100 probe
UT-ONE S16B Gold-plated banana conectors
UT-ONE S16B with T type Thermocouple
UT-ONE S16B with Pt100 probes and Ambient conditions probe
UT-ONE S16B  with a 10 kΩ Thermistor

UT-ONE Applications Your browser does not support SVG.
Ambient Conditions Probe Your browser does not support SVG.
Ambient conditions probe
Ambient conditions probe with 2 m extension cable
Ambient conditions probe connector without plastic protective plug
Ambient conditions probe - direct connection
Ambient conditions probe - connection with 2 m extension cable

Ambient probe is used for the measurement of ambient conditions (air temperature - TA and relative humidity - RH). It should be connected to the 4-pole circular connector located above the USB and RS232 sockets, directly or indirectly with the 2m extension cable.

Specifications for TA probe
Range:
-20 °C to 70 °C
Accuracy:
±0.5 °C
Resolution:
0.01 °C
Sampling period:
4 seconds
Sensor type:
Sensirion SHT75
Specifications for RH probe
Range:
10 % to 90 % (non condensing)
Accuracy:
±2%
Resolution:
0.1 %
Sampling period:
4 seconds
Sensor type:
Sensirion SHT75

To get a quote or more information contact us at info@batemika.com .